JINLING-840 Anesthesia Unit

Specifications

   Flow rate: O2 0.05 ~ 1 L/min; 1~ 10 L/min   

N2O 0.05 ~ 1 L/min; 1 ~10 L/min

                        Air 0.2 ~ 1 L/min; 1 ~ 12 L/min  (ORC O2 ≥ 25%)

■ O2 flush: 35 ~ 75 L/min

■ Insufficient oxygen alarm and automatic N2Ocalibration function:

When O2 level is ≤ 0.2Mpa, audible alarm sounds and lasts for at least 7 seconds

When O2 level is ≤ 0.14Mpa, N2O flow decreases correspondingly

When O2 level is ≤ 0.04Mpa, N2O flow is cut off completely

■ Integrated breathing circuit system, APL adjustment range: 0.5~ 7 kPa

■ CO2 absorber capacity: 1.6 L

■ Vaporizer ambient temperature: 15 ~ 35 °C

■ Input flow: 0.25 ~ 15 L/min

■ Output back pressure: -10 kPa ~ 20 kPa

■ Output concentration: 0.2 % ~ 5% (± 0.2 %) (Vol %)

■ Ventilation frequency: 6 ~ 40 bpm

■ Tidal volume; adult: 100 ~ 1100 ml Infant: 30 ~ 150 ml

■ I:E: ratio 1:1 ~ 1:9.9 (adjustable)